IBS Institute for Basic Science
Search

Closed Cycle Refrigerator System (Old CCR)

equipment explanation
Model CCR-1020
Operating time SUN(00:00~24:00)
MON(00:00~24:00)
TUE(00:00~24:00)
WED(00:00~24:00)
THU(00:00~24:00)
FRI(00:00~24:00)
SAT(00:00~24:00)
Location 86691
inquiry Hojoon Yi
clainel01@gmail.com
Reservation

Notice


16.05.16 Notice has renewed.


Closed cycle refrigerator system consists of chamber cooling system and feedthrough which have connection with device inner chamber and outer port.

By using this system, we can measure electrical properties in 60K to 350K.

 

Machine Status:

Vacuum level : 10-6 Torr

Temperature Range : 60K to 350K

Sample Loading :   Sample needs to be mounted on designated socket before it can be measured (Please contact super-user for further information)

Measurement Scheme : All measurements using this machine are performed using Keithley instruments (which are programmed using labview).

Sample Connection : The sample is connected to external instruments using BNC Cables